Whitley H35 Anaerobic Workstation

 

 

Das Hans-Knöll-Institut aus Jena, Institut für Naturstoff-Forschung und Infektionsbiologie, hat sich im Rahmen Ihrer Arbeit mit der Gruppe Biomolekulare Chemie, unter der Leitung von Prof. Christian Hertweck, mit der Untersuchung von Naturstoffen aus anaeroben Bakterien befasst. Im Rahmen dieser Arbeit wurde die Whitley A35 dabei als anaerobe Arbeitsstation getestet.

 

Die Erfahrungen dieser Arbeit wurde uns von Herrn Florian Bredy als Anwenderbericht zur Verfügung gestellt. Um den kompletten Bericht sehen zu können, klicken Sie HIER!

 

Did you know: How best to clean your workstation?

Laboratory Standard Shot DYK Blog

 

 

This is the 10th in our ‘Did you know’ series of articles introducing things you may not know about the Whitley range of products and services – read on and you may find out something about your workstation that could make your working life easier.

We regularly receive enquiries from customers with regard to which disinfectants can be safely used inside Whitley Workstations. Although it is not practicable to evaluate every available product, we have tested the materials compatibility of several sporicidal disinfectants and can now provide comprehensive advice on this topic.

 

 

How often should I clean my workstation?

DWS recommend that as part of the daily checks, users ensure that the workstation is free from spillage and cleaning carried out as necessary. Prior to having your workstation serviced, however, it may be necessary for it to be decontaminated.

Which disinfectants have DWS tested?

  • Spor-Klenz – peracetic acid (approx 0.1%) + hydrogen peroxide (approx 1.0%) – “ready to use” product, tested without further dilution.
  • Chemgene HLD4H – quaternary ammonium compounds (approx 1%) + chlorhexidine digluconate (approx 1%) – “ready to use” product, tested without further dilution.
  • Safe Hospital – hypochlorous acid – “ready to use” product, tested without further dilution.
  • Tristel Fuse – chlorine dioxide (approx 1%) – prepared from concentrated components according to manufacturer’s instructions.

For more information on these tests, please see the appendices of your Whitley Workstation user manual.

 

Which of the above products are OK for use in a Whitley Workstation?

Our tests demonstrated that each of these formulations is satisfactory for use in Whitley Workstations and will not cause any damage when used in accordance with the manufacturer’s recommendations.

 

What if I want to use a cleaning agent not on the above list?

Other disinfectants containing the active ingredients listed above, at similar concentrations to those shown, can also be safely used inside Whitley Workstations. If you wish to use a disinfectant containing active ingredients not listed here, please contact us for advice before proceeding.

 

Why should I contact DWS for advice – will any damage not be covered under my warranty?

Unfortunately, we cannot cover under the equipment’s warranty any damage caused to a Whitley Workstation as a result of exposure to products that have not been approved by us.

 

Can I use Virkon as my preferred sanitising solution?

After a thorough study lasting 20 weeks, we concluded that the presence of any uncovered containers of Virkon in anaerobic and variable atmosphere workstations leads to the degradation of stainless steel, brass and some other metal components. The effect is more considerable when oxygen is present in the selected gas mixture. These conclusions are supported by technical information available on the Virkon manufacturer’s website and independent observations by an Institute of Materials metallurgist. If Virkon is the preferred sanitising solution these adverse effects can be minimised by ensuring that Virkon within any workstation is always kept in a covered vessel when not in use.

 

OK, so what’s the best way to clean the acrylic parts of a Whitley Workstation?Labdet100 Blog

The transparent and/or white acrylic on the inside and outside of the Whitley Workstation System may be swabbed with a 2% solution of Labdet 100 (DWS – stock code D00003) in warm water and dried afterwards with a soft clean cloth. In the case of culture spillages then any of the disinfectants listed above solution should be applied to the spillage and left for 30 minutes. It should then be mopped.

 

Never use any solvent on the acrylic surfaces of the workstation. Use only water and a mild detergent solution (i.e. Labdet 100 2% solution) as a cleaning agent.

 

What if the spillage leaks underneath the floor of the workstation?

If spillage is not contained on the working surface then access underneath this area will be required in all Whitley Workstations except the DG250. Some workstations require the removal of retaining bolts before the floor can be lifted up or pivoted for cleaning underneath.

If liquid has been spilled upon the Anotox and catalyst sachets they should be removed, dried and/or replaced.

 

How can I deal with scratches on the acrylic plastic surface of my workstation?

Scratches may be removed by gently polishing the surface with ‘DURAGLIT WADDING’ followed by wiping with a soft clean cloth. Deep scratches may require the use of Wet and Dry abrasive paper used wet, followed by polishing with ‘DURAGLIT’ – seek advice from Don Whitley Scientific Limited or our authorised agents overseas.

 

How can I clean the sleeves on my workstation?Sleeve Kit-Components DYK Blog

The white sleeves used on all Whitley Workstations may require cleaning after periods of extended use. There are a number of options for cleaning and some recommendations listed below:

  • Wipe with a mild detergent
  • Wash or dry-clean with a detergent up to 75ºC
  • For disinfection use Sodium Hypochlorite (at no more than 1000ppm active Chlorine) or any of the disinfectants listed above
  • Do not use Biological or Phenolic based cleaners
  • Wash and dry completely before storage
  • Some surface wrinkling may occur.

 

If you have any remaining concerns about disinfectants for use in Whitley Workstations, please get in touch with our sales team:

Tel: 01274 595728

Or

Email: sales@meintrup-dws.de

 

Practical Guide to Clinical Anaerobic Bacteriology

Introduction to Clinical Anaerobic Bacteriology

Eighteen months ago, Don Whitley Scientific played an integral role in producing An Introduction to Clinical Anaerobic Bacteriology. This unique guide is being very well-received by scientists in laboratories nationwide. It enables the reader to isolate and identify 12 commonly occurring clinically important anaerobic bacteria. It contains 60 pages printed on waterproof, tear-resistant synthetic paper to withstand regular use at the bench and is illustrated with stunningly detailed colour photographs and attractive reference tables.

Designed to fill a gap in the practical reference materials currently available to support clinical laboratory practice for microbiologists, the publication was written by Professor Michael W D Wren, MBE FIBMS, former consultant Biomedical Scientist in the microbiology department, University College Hospital, and visiting Professor at The University of Westminster. With a foreword by Professor Brian Duerden, Emeritus Professor of Medical Microbiology at Cardiff University and input from Dr Don Whitley, Chairman and founder of Don Whitley Scientific Limited, considerable expert knowledge has been combined to make this the most up-to-date reference guide for the modern clinical laboratory.

Images were kindly supplied by the Anaerobe Reference Unit, Cardiff.

For further information, or to order your copy, please contact us at sales@meintrup-dws.de or call us on 01274 595728.

 

Educational Day at JCB

Under a digger

26 hardy DWS employees got up before dawn on 10th October to take a trip to JCB in Rocester. JCB is a family-owned company, like Don Whitley Scientific Ltd, and we share many of the same values. The visit was organised to give us some insight into a successful, global company and hopefully pick up some ideas for our own future planning and growth strategy.